Chapter 2
49
Performance Tests
Individual Test Description
2. Pe
rfor
ma
nce T
e
st
Step 3.
Preset the Signal Generator. Then set the Signal Generator as follows.
Step 4.
Press
[Trigger]
-
Single
of E5052A to make a measurement.
Step 5.
Record the Phase Noise in the Performance Test Record.
Table 2-13
E5052A Settings Value (for Option 011)
E5052A settings
Key strokes
Trigger Mode : Hold
[Trigger]
-
Trigger to Phase Noise
-
Hold
Frequency Band : 39 M - 101
MHz
[Setup]
-
Frequency Band
-
39M-101MHz
Start Frequency : 100 Hz
[Start/Center]
-
100Hz
Stop Frequency : 10 MHz
[Stop/Span]
-
10MHz
Attenuator : 0 dB
[Attn]
-
Input Attenuator
-
[0]
-
[x1]
Maker 1’s Position : 1 MHz
[Marker]
-
Marker 1
-
[1]
-
[M/u]
Maker 2’s Position : 3 MHz
[Marker]
-
Marker 2
-
[3]
-
[M/u]
Maker 3’s Position : 7 MHz
[Marker]
-
Marker 3
-
[7]
-
[M/u]
Instrument
Setting
Signal Generator
Frequency : 70 MHz
Power : 8 dBm , RF Output : ON
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...