Appendix B
211
System Recovery
System Recovery
B.
Sy
stem
Recovery
Procedure to execute the user recovery function
Returns the contents of the C drive to a user-specified state. To use this function, you must
create the user backup image in advance. For more information, see the description
“Procedure to create the user backup image” on page 208.
NOTE
This function is available when the volume label on the hard disk is AL200 or higher.
NOTE
You need the keyboard for this operation.
Step 1.
Shut down the E5061A/E5062A.
Step 2.
Connect the keyboard to the E5061A/E5062A.
Step 3.
Insert the disk for the system recovery into the floppy disk drive of the E5061A/E5062A.
Step 4.
Press the standby switch of the E5061A/E5062A to turn it on.
Step 5.
When the screen as shown in the figure below appears, press and hold
of the
keyboard until this screen disappears.
NOTE
After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Summary of Contents for E5061A
Page 142: ...142 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 9 Front Panel ...
Page 168: ...168 Chapter6 Replacement Procedure Front Panel Removal Figure 6 2 Front Panel Removal ...
Page 192: ...192 Chapter6 Replacement Procedure A51 LCD Display Replacement Figure 6 16 A51 LCD Removal ...
Page 194: ...194 Chapter6 Replacement Procedure LCD Backlight Replacement ...
Page 198: ...198 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 202: ...202 AppendixA Manual Changes Manual Changes ...
Page 218: ...218 AppendixB System Recovery Calibration of the Touch Screen ...
Page 233: ...233 E Power Requirement E Power Requirement ...