Chapter 2
37
Performance Tests
E5061A Performance Test Record (for Option 150/250)
2. Pe
rfor
ma
nce T
e
st
E5061A Performance Test Record (for Option 150/250)
Agilent Technologies E5061A RF Network Analyzer
Frequency Accuracy Test
RF Output Level Accuracy and Flatness Test
Level Accuracy Test (@ Port 1)
Level Flatness Test (@ Port 1) (at 0dBm, relative to 50 MHz reference)
Serial Number:
Option:
Temperature:
C
Test Date:
Humidity:
% R.H.
Tested by:
Frequency [Hz]
Test limit [Hz]
Test result [Hz]
Measurement
uncertainty [Hz]
300 k
1.5
0.056
50 M
250
9
1.5 G
7.5 k
k
0.28 k
Power level
[dBm]
Frequency
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
0
50 M
0.8
0.07
Frequency [Hz]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
300 k
1.0
0.27
1.5 M
1.0
0.07
3.00 M
1.0
0.07
7.00 M
1.0
0.07
15.0 M
1.0
0.07
275 M
1.0
0.07
575 M
1.0
0.07
875 M
1.0
0.07
1.175 G
1.0
0.07
Summary of Contents for E5061A
Page 142: ...142 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 9 Front Panel ...
Page 168: ...168 Chapter6 Replacement Procedure Front Panel Removal Figure 6 2 Front Panel Removal ...
Page 192: ...192 Chapter6 Replacement Procedure A51 LCD Display Replacement Figure 6 16 A51 LCD Removal ...
Page 194: ...194 Chapter6 Replacement Procedure LCD Backlight Replacement ...
Page 198: ...198 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 202: ...202 AppendixA Manual Changes Manual Changes ...
Page 218: ...218 AppendixB System Recovery Calibration of the Touch Screen ...
Page 233: ...233 E Power Requirement E Power Requirement ...