Chapter 2
65
Performance Tests
E5062A Performance Test Record (for Option 150/250)
2. Pe
rfor
ma
nce T
e
st
Dynamic Accuracy Test
@ Reference power level: -10 dBm
Direction: S21
Direction: S12 (Option 250 only)
Power level [dBm]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
10
0.303
0.011
5
0.144
0.011
0
0.090
0.011
-5
0.068
0.011
-15
0.065
0.011
-20
0.073
0.011
-25
0.080
0.015
-30
0.087
0.015
-35
0.095
0.019
-40
0.103
0.019
-45
0.111
0.022
-50
0.121
0.022
-55
0.134
0.024
-60
0.150
0.024
-65
0.174
0.026
-70
0.211
0.026
-75
0.270
0.028
-80
0.371
0.028
-85
0.546
0.030
-90
0.841
0.030
-95
1.334
0.032
-100
2.141
0.032
Power level [dBm]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
10
0.303
0.011
5
0.144
0.011
Summary of Contents for E5061A
Page 142: ...142 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 9 Front Panel ...
Page 168: ...168 Chapter6 Replacement Procedure Front Panel Removal Figure 6 2 Front Panel Removal ...
Page 192: ...192 Chapter6 Replacement Procedure A51 LCD Display Replacement Figure 6 16 A51 LCD Removal ...
Page 194: ...194 Chapter6 Replacement Procedure LCD Backlight Replacement ...
Page 198: ...198 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 202: ...202 AppendixA Manual Changes Manual Changes ...
Page 218: ...218 AppendixB System Recovery Calibration of the Touch Screen ...
Page 233: ...233 E Power Requirement E Power Requirement ...