42
Chapter 2
Troubleshooting
Performance Test failure Troubleshooting
Performance Test failure Troubleshooting
This section describes the adjustment and troubleshooting procedures used when the
E5061B fails the performance tests. If the performance of the instrument is critical for the
test limits and seems to be adjustable, perform first the adjustment(s) related to the failed
test. When the test result is far from the tolerance of the test or the performance is not
adjustable, isolate the faulty assembly in accordance with the "Performance Tests failure
Troubleshooting procedure".
Recommended adjustment for Performance Test failure
Table 2-6 and Table 2-7 shows the recommended adjustments when the performance test
fails. Select the adjustment program corresponding to the recommended adjustment and
perform the adjustment.
Table 2-6
Recommended adjustment for performance test failure (Option 3L5)
Te
st
N
u
m
.
First failed test
Recommended adjustment
Oven r
efer
ence
Fr
equency r
efer
ence
Loca
l gain
Mixer local leakage
Receiver absolute gain Rch
Synthes
izer gain
Sou
rce out
put
p
ower
DC bias
Receiver IF
ra
nge
Receiver port characteristics
Receiver absolute gain T
ch
Receiver absolute gain L
F
Sour
ce output power (LF)
DC lev
el mo
ni
tor
1
Frequency accuracy
test
√
2
Frequency accuracy
test (Opt. 1E5)
√
3
RF output level
accuracy and
flatness test
√
√
√
√
4
RF output level
linearity test
√
√
√
√
5
Trace noise test
6
Crosstalk & system
dynamic range test
√
√
7
Dynamic accuracy
test
√
√
8
Uncorrected system
performance test
√
√
Summary of Contents for E5061B
Page 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Page 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Page 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Page 116: ...115 E Power Requirement D Power Requirement ...
Page 123: ...122 AppendixE Messages ...