Page 2 of 5
E5062A-04A
Situation:
A very few of clock signal generation component in the receiver board might not work properly and as
a result, the noise floor increases although the IFBW is set to narrower bandwidth.
The typical phenomenon is that the system dynamic range exceeds the following specification right
after the E5062A power is turned on, especially in a low temperature environment.
System dynamic range
Specification
Supplemental information
300 kHz to 3 GHz, IF bandwidth = 3 kHz
90 dB
95 dB
300 kHz to 3 GHz, IF bandwidth = 10 Hz
115 dB
120 dB
Solution/Action:
Replace A2 receiver module.
(For 50 Ohm Option; Opt. 150 and Opt. 250)
P/N
Description
Qty
E5062-69011 A2 Receiver 50 Ohm (Exchange)
1
E5062-62011 A2 Receiver 50 Ohm
1
(For 75 Ohm Option; Opt. 175 and Opt. 275)
P/N
Description
Qty
E5062-69012
A2 Receiver 75 Ohm (Exchange)
1
E5062-62012 A2 Receiver 75 Ohm
1