102
Chapter 4
Calibration
Checking Calibration Status
Checking Calibration Status
Execution status of error correction for each channel
You can check the execution status of error correction for each channel with the error
correction status.
The error correction status is indicated in the channel status bar in the lower part of the
window by the symbols in the below table.
Symbol
Execution status of error correction
Cor
(displayed in blue)
Error correction: On (enabled for all traces)
Cor
(displayed in gray)
Error correction: On (enabled for some traces)
Off
(displayed in gray)
Error correction: Off
---
(displayed in gray)
Error correction: On (no calibration data)
C?
(displayed in blue)
Error correction: On (Interpolation is being executed or the IF
bandwidth, power level, power range, sweep time, sweep delay time,
sweep mode, or sweep type is different from that when the calibration
was executed.)
C!
(displayed in blue)
Error correction: On (Extrapolation is being executed.)
Execution status of error correction for each trace
You can check the status of the error correction actually executed for each trace with the
trace status area.
For a trace for which error correction is executed, the applied calibration type is indicated
in the trace status area by the symbols in the table below.
Symbol
Calibration type
RO
Open response calibration
RS
Short response calibration
RT
Thru response calibration
ER
Enhanced response calibration
F1
1-port calibration
F2
Full 2-port calibration/2-port TRL calibration
F3
Full 3-port calibration/3-port TRL calibration
*1.The simplified full 3/4-port calibration and the simplified 3/4-port TRL calibration (that
acquire the calibration coefficients skipping part of measurement) are not discriminated
from the normal full 3/4-port calibration and 3/4-port TRL calibration, and
F3
or
F4
is
displayed.
*1
F4
Full 4-port calibration/4-port TRL calibration
If none of the symbols described above is displayed, error correction is not executed for the
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...