106
Chapter 4
Calibration
Setting the trigger source for calibration
Setting the trigger source for calibration
You can set the trigger source for calibration before executing calibration.
You can select it from “Internal” or “System.” Setting it to “System” allows you to use the
same trigger source setting for calibration and measurement.
You can set the trigger source for calibration (which is usually set to “Internal”) to
“System,” which allows you to generate triggers at any timing from a PC for external
control or from the front panel for calibration in the same way as for measurement.
The setting of the point trigger and averaging trigger is also applied to the trigger for
calibration. When the trigger source for calibration is set to “System” and the trigger
source for measurement is set to “External” or “Manual” with the point trigger function set
to ON, a trigger is required for each measurement point during calibration.When the
averaging trigger function is set to ON, the sweep is performed the number of times
specified by the averaging factor for a single trigger during calibration.
NOTE
For the following types of calibration, the setting of the trigger source does not take effect.
Those calibration operations are controlled by the internal trigger.
•
Calibration using ECal
•
Power calibration
•
Receiver calibration
•
Mixer converter calibration
Follow these steps to set the trigger source for calibration.
Step 1.
Press
.
Step 2.
Press
Cal Trig Source
.
Step 3.
Select the trigger source you want to use.
Softkey
Function
Internal
Selects “Internal.”
System
Selects “System.”
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...