Chapter 4
107
Calibration
OPEN/SHORT Response Calibration (reflection test)
4. Cal
ibra
tion
OPEN/SHORT Response Calibration (reflection test)
In OPEN or SHORT response calibration, calibration data are measured by connecting an
OPEN or SHORT standard, respectively, to the desired test port. For frequency response,
these calibrations effectively eliminate the reflection tracking error from the test setup in a
reflection test using that port (Figure 4-3). It is also possible to carry out isolation
calibration with a LOAD standard during OPEN/SHORT response calibration. An
isolation calibration will eliminate the directivity error from the test setup in a reflection
test using that port (Figure 4-4).
Figure 4-3
1-Port error model (OPEN/SHORT response)
Figure 4-4
1-Port error model (OPEN/SHORT re isolation)
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...