Chapter 4
117
Calibration
Enhanced Response Calibration
4. Cal
ibra
tion
NOTE
An asterisk (*) in the upper right of the softkey indicates that, if you select the test port and
execute acquisition of the calibration coefficient (press
Done
), the existing calibration
coefficient will be cleared.
Step 6.
Connect an OPEN calibration standard to the output port.
Step 7.
Press
Open
to start the calibration measurement.
Step 8.
Disconnect the OPEN calibration standard that was connected in Step 8 and replace it with
a SHORT calibration standard.
Step 9.
Press
Short
to start the calibration measurement.
Step 10.
Disconnect the SHORT calibration standard that was connected in Step 8 and replace it
with a LOAD standard.
Step 11.
Press
Load
to start the calibration measurement.
Step 12.
Make a THRU connection between the two ports.
Step 13.
Press
Thru
to start the calibration measurement.
Step 14.
If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a.
Connect a LOAD standard to the two test ports.
b.
Press
Isolation (Optional)
to start the calibration measurement.
Step 15.
Press
Return
.
Step 16.
Press
Done
to terminate the enhanced response calibration process. Upon pressing the key,
calibration coefficients will be calculated and saved. The error correction function will also
be automatically enabled.
1-4(S14 S44)
Measure between test ports 4 and 1. selecting port 4 as the output.
2-4(S24 S44)
Measure between test ports 4 and 2. selecting port 4 as the output.
3-4(S34 S44)
Measure between test ports 4 and 3. selecting port 4 as the output.
Softkey display
Function
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...