122
Chapter 4
Calibration
Full 3-Port Calibration
Figure 4-18
Connecting standards in full 3-port calibration
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
3-Port Cal
.
Step 5.
Press
Select Ports
.
Step 6.
Select the test ports on which you will perform full 3-port calibration. (In the procedure
below, the selected test ports are denoted as x, y, and z.)
Softkey
Function
1-2-3
Selects test ports 1, 2, and 3
1-2-4
Selects test ports 1, 2, and 4
1-3-4
Selects test ports 1, 3, and 4
2-3-4
Selects test ports 2, 3, and 4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the