Chapter 4
Calibration
ECal (electronic calibration)
Step 3.
Press
or
to select the channel for which you want to perform the
calibration.
Step 4.
Press
.
Step 5.
Press
ECal
.
Step 6.
Press
1 Port ECal
.
Step 7.
Perform a 1-port calibration.
Softkey
Function
Port 1
Performs a 1-port calibration on test port 1
Port 2
Performs a 1-port calibration on test port 2
Port 3
*1.Only with Options 313, 314, 413, and 414.
*1
Performs a 1-port calibration on test port 3
Port 4
*2.Only with Options 413 and 414.
*2
Performs a 1-port calibration on test port 4
Full 2-Port Calibration Using the 2-Port ECal Module
Follow the procedure below to perform a full 2-port calibration using the 2-port ECal
module.
Step 1.
Connect the USB port on the ECal module with the USB port on the E5070B/E5071B via a
USB cable. This connection may be done while the E5070B/E5071B’s power is on.
Step 2.
Connect port A and port B on the ECal module to the test ports to be calibrated.
NOTE
You can connect the ports of the ECal and the test ports of the E5070B/E5071B
arbitrarily.Connected ports can be manually specified although they are automatically
detected before data measurement.
For more information, “Turning off ECal auto-detect
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...