Chapter 4
129
Calibration
ECal (electronic calibration)
4. Cal
ibra
tion
Figure 4-22
Connecting ECal module (full 2-port calibration)
Step 3.
Press
or
to select the channel for which you want to perform the
calibration.
Step 4.
Press
.
Step 5.
Press
ECal
.
Step 6.
To enable isolation calibration, press
Isolation
and confirm that the display has turned
ON
.
Step 7.
Press
2 Port ECal
. When using a 2-port E5070B/E5071B (Options 213 or 214), pressing
this key performs a 2-port ECal.
Step 8.
When using a 3-port or 4-port E5070B/E5071B (Options 313, 314, 413, or 414), press one
of the softkeys below to start a full 2-port calibration.
Softkey
Function
Port 1-2
Performs a full 2-port calibration between test ports 1 and 2
Port 1-3
Performs a full 2-port calibration between test ports 1 and 3
Port 1-4
*1.Only with Options 413 and 414.
*1
Performs a full 2-port calibration between test ports 1 and 4
Port 2-3
Performs a full 2-port calibration between test ports 2 and 3
Port 2-4
Performs a full 2-port calibration between test ports 2 and 4
Port 3-4
Performs a full 2-port calibration between test ports 3 and 4
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...