Calibration
ECal (electronic calibration)
4. Cal
ibra
tion
Turning off ECal auto-detect function
The ECal module automatically detects the connection between E5070B/E5071B's test
ports and ECal module's ports. You can turn off this function to set ports manually.
NOTE
Even if the connection is wrong with the auto-detect function turned off, no error is
displayed.
Step 1.
Press
or
to select the channel for which you want to turn off the
auto-detect function.
Step 2.
Press
.
Step 3.
Press
ECal
.
Step 4.
Press
Orientation
, and select
Manual
.
Softkey
Function
Auto
Turns on the auto-detect function.
Manual
Turns off the auto-detect function.
Step 5.
Specify a test port of the E5070B/E5071B.
Softkey
Function
Port 1
Specifies port 1 of the E5070B/E5071B.
Port 2
Specifies port 2 of the E5070B/E5071B.
Port 3
*1.Only with Options 313, 314, 413, and 414.
*1
Specifies port 3 of the E5070B/E5071B.
Port 4
*2.Only with Options 413 and 414.
*2
Specifies port 4 of the E5070B/E5071B.
Step 6.
Specify an Ecal port for the port of the E5070B/E5071B you specified.
Softkey
Function
Port A
Specifies port A of Ecal.
Port B
Specifies port B of Ecal.
Port C
*1.Only with Ecal modules supporting 3 ports and 4 ports.
*1
Specifies port C of Ecal.
Port D
*2.Only with Ecal modules supporting 4 ports.
*2
Specifies port D of Ecal.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...