138
Chapter 4
Calibration
2-port TRL calibration
Operational procedure
Figure 4-31
Connection of standards in 2-port TRL calibration
NOTE
The shapes of actual standards differ from those shown in Figure 4-31 which are just
symbols for illustration.
Step 1.
Press
or
to select the channel for which you want to execute
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
2-Port TRL Cal
.
Step 5.
Press
Select Ports
.
Step 6.
Select the test ports for which you want to execute TRL 2-port calibration. (In the
following procedure, test ports you select here are expressed as test ports x and y.)
Softkey
Function
1-2
Selects test ports 1 and 2.
1-3
Selects test ports 1 and 3.
1-4
Selects test ports 1 and 4.
2-3
Selects test ports 2 and 3.
2-4
Selects test ports 2 and 4.
3-4
Selects test ports 3 and 4.
NOTE
An asterisk (*) in the upper right of the softkey indicates that a calibration coefficient exists
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...