162
Chapter 4
Calibration
Adapter Characterization
How to execute adapter characterization
1. Setting Stimulus Conditions
Set the stimulus conditions of the channel for which you will execute the adapter
characterization.
For information on the setting procedure, see Chapter 3, “Setting Measurement
Conditions.”
2. Performing Calibration
Perform 1-port calibration for test port 1 and test port 2 in the channel for which the
stimulus condition has been set. Use the connector type appropriate for the adapter for the
calibration surface of test port 1 and test port 2.
For information on the procedure for performing this calibration, see “1-Port Calibration
(reflection test)” on page 113.
3. Starting VBA Macro
Step 1.
Press
.
Step 2.
Press
Load Project
.
Step 3.
The Open dialog box appears. Specify the file name
“D:\Agilent\AdapterCharacterization.vba” and press the
Open
button.
Step 4.
Press
to start the macro.
Step 5.
Select
Characterize Adapter
(1 in Figure 4-41) to display the Adapter Characterization
screen.
Figure 4-41
Adapter Characterization macro (Adapter Characterization screen)
4. Selecting Channel
Select the channel for which calibration has been performed (2 in Figure 4-41).
NOTE
The selected channel has no relation to the active channel.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...