170
Chapter 4
Calibration
Adapter Removal-Insertion
Procedure for Adapter Removal / Insertion
The s-parameter of a reciprocal adapter can be determined when the following data is
available.
•
Open, Short, and Load measurements.
•
Actual values derived from the CalKit definitions.
•
An approximate length of the adapter.
•
Nature of the intended operation : removal or insertion.
Once the s-parameter has been determined, run de-embedding and update the error
coefficients.
NOTE
If the frequency setting is not zero span, an approximate length of the adapter should be
provided otherwise an error generated in the measurement.
Step 1.
Perform a full n-port calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
Adapter Removal
.
Step 5.
Select the port in/from which you want to insert/remove adapter characteristics from
Select Port
.
Step 6.
Press
Cal Kit
, and then select calibration kit.
Step 7.
Select the adapter length.
Softkey
Function
AUTO
Adapter automatic
Positive Value
Adapter removal
Negative Value
Adapter insertion
NOTE
For adapter removal, the adapter length is positive . For adapter insertion, the adapter
length is negative.
Step 8.
Press
Open
,
Short
, and
Load
respectively.
Step 9.
Press
Done
to complete the process.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...