Chapter 4
191
Calibration
Changing the Calibration Kit Definition
4. Cal
ibra
tion
Setting a media type for the calibration kit
You can set a media type for the standard you use.
Setting procedure
Step 1.
Press
.
Step 2.
Press
Cal Kit
, and select a calibration kit.
Step 3.
Press
Modify Cal Kit
.
Step 4.
Press
Define STDs
, and select a standard.
Step 5.
Press
Media
, and select a media type.
Softkey
Function
Coaxial
Selects coaxial as the media type.
Waveguide
Selects waveguide as the media type.
NOTE
If you select waveguide as the media type, set the system impedance and the characteristic
impedance to 1
Ω
.
NOTE
Depending on the media type you use, the calculation method of the electrical delay, which
is required to correct the phase delay, differs. For more information, see “Electrical Delay”
on page 398.