196
Chapter 4
Calibration
Changing the Calibration Kit Definition
85032B
1. Short(m)
2. Open(m)
3. Broadband
4. Thru
Label
Short(m)
Open(m)
Broadband
Thru
STD Type
Short
Open
Load
Delay/Thru
C0 [
×
10
-15
F]
0
119.09
0
0
C1 [
×
10
-27
F/Hz]
0
-36.955
0
0
C2 [
×
10
-36
F/Hz
2
]
0
26.258
0
0
C3 [
×
10
-45
F/Hz
3
]
0
5.5136
0
0
L0 [
×
10
-12
H]
0
0
0
0
L1 [
×
10
-24
H/Hz]
0
0
0
0
L2 [
×
10
-33
H/Hz
2
]
0
0
0
0
L3 [
×
10
-42
H/Hz
3
]
0
0
0
0
Offset Delay [s]
93 f
0
0
0
Offset Z0 [
Ω
]
49.992
50
50
50
Offset Loss [
Ω
/s]
700 M
700 M
700 M
700 M
Arb. Impedance [
Ω
]
50
50
50
50
7. Short(f)
8. Open(f)
Label
Short(f)
Open(f)
STD Type
Short
Open
C0 [
×
10
-15
F]
0
62.14
C1 [
×
10
-27
F/Hz]
0
-143.07
C2 [
×
10
-36
F/Hz
2
]
0
82.92
C3 [
×
10
-45
F/Hz
3
]
0
0.76
L0 [
×
10
-12
H]
0
0
L1 [
×
10
-24
H/Hz]
0
0
L2 [
×
10
-33
H/Hz
2
]
0
0
L3 [
×
10
-42
H/Hz
3
]
0
0
Offset Delay [s]
17.817 p
17.411 p
Offset Z0 [
Ω
]
50.209
50
Offset Loss [
Ω
/s]
2.1002 G
700 M
Arb. Impedance [
Ω
]
50
50
NOTE
The designations (m) and (f) in the name (label) of the standard indicate male (m) and
female (f) analyzer connectors.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...