242
Chapter 5
Making Measurements
Setting Up the Trigger and Making Measurements
Setting Up the Trigger and Making Measurements
The E5070B/E5071B has one trigger source. When this trigger source detects a trigger
signal that has occurred, a sweep is performed for channels in the “Initiate” state in the
order of channel 1 to channel 16. You set the “Initiate” or “Idle” status of each channel by
changing the trigger mode. For details on the trigger system, refer to
Programmer's Guide
.
NOTE
The execution of measurement for each channel does not depend on whether the channel is
displayed. Channels that have been activated can be measured even if they are not
displayed.
For each channel, a sweep is performed only for the stimulus ports needed to update the
parameters of the displayed trace.
Sweep Order in Each Channel
In a channel, each test port is set to a stimulus port in the order of port number and updates
each trace (Table 5-1).
Table 5-1
Sweep Order in Each Channel
Sweep Order
Stimulus Port
Updated Trace
Port 1
S11, S21, S31
, S41
Port 2
S12, S22, S32
Port 3
*1.Option 313, 314, 413, or 414 only
*1
S13
, S23
Port 4
*2.Option 413 or 414 only
*2
S14
, S24
NOTE
If full 2-, 3-, or 4-port error correction is in effect, no trace between calibrated ports is
updated until the last calibrated port is swept as the stimulus port.
Sweep is not executed for stimulus ports that are not required for updating traces.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...