256
Chapter 6
Data Analysis
Analyzing Data on the Trace Using the Marker
Reading the Relative Value From the Reference Point on the Trace
You can convert the marker reading into a relative value from the reference point.
Figure 6-1
Delta marker mode
Converting From a Reference Point to a Relative Value
Step 1.
Following Step 1 to Step 5 in “Reading values on the trace” on page 253, place the
reference marker on the point to be used as the reference.
Step 2.
Press
Ref Marker Mode
to turn on the reference mode.
With the reference mode turned on, the stimulus values and response values are indicated
in relative values referred to by the position of the reference marker.
Step 3.
Following Step 4 to Step 5 in “Reading values on the trace” on page 253, place markers 1
through 9 on the desired points to read the values.
NOTE
Pressing
Marker
→
Ref Marker
enables you to place the reference marker at the position of
the currently active marker. The reference mode will then turn on automatically.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...