Chapter 7
299
Fixture Simulator
Evaluating Balanced Devices (balance-unbalance conversion function)
7
. F
ixt
ur
e
Sim
ula
to
r
Step 7.
Press
Return
.
Step 8.
Press
BalUn
to turn on the balanced/unbalanced state conversion function.
Step 9.
Press
Fixture Simulator
to turn on the fixture simulator function.
Steps for Measurement Parameter Setups
Performing balance-unbalance conversion enables you to make measurements with mixed
mode S-parameters, imbalance parameters and CMRR. Parameters that can be used differ
depending on the balance-unbalance topology specified in Step 5 in “Steps for
Balance-Unbalance Conversion” on page 298.
Step 1.
Press
(or
-
Fixture Simulator
-
Measurement
).
Step 2.
Select the measurement parameter.
Checking device type and port assignment
You can check the device type and the port assignment for the balance-unbalance
conversion by displaying the balance measurement topology property as shown in Figure
7-16.
Figure 7-16
Balance measurement topology property
Procedure to Turn On/Off Balance Measurement Topology Property Display
Follow these steps to turn on/off the balance measurement topology property display.
Step 1.
Press
.
Step 2.
Press
Fixture Simulator
.
Step 3.
Press
Topology
.
Step 4.
Press
Property
. Each press toggles between on/off.
Port 2 (se)
Selects a port on the analyzer from among
1
,
2
,
3
, and
4
for connection to
logical 2 (Port b in Figure 7-8)
Port 3 (bal)
Selects two ports on the analyzer from among
1-2
,
1-3
,
1-4
,
2-1
,
2-3
,
2-4
,
3-1
,
3-2
,
3-4
,
4-1
,
4-2
, and
4-3
for connection to logical port 3 (Port c and Port d in
Softkey
Function
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...