322
Chapter 8
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
Step 3.
Use the
X-Axis
key to set the frequency data.
Softkey
Function
Normal
Sets the basic frequency as frequency data
RF+LO
Sets IF frequency (RF+LO) as frequency data
RF-LO
Sets IF frequency (RF-LO) as frequency data
LO-RF
Sets IF frequency (LO-RF) as frequency data
Base
Sets the basic frequency as frequency data
Stimulus
Sets the stimulus port frequency as frequency data
Response
Sets the response port frequency as frequency data
NOTE
When
Frequency Offset
is
OFF
, and
LO Frequency
is
ON
(mixer measurement by
vector-mixer calibration),
Normal
,
RF+LO
,
RF-LO
and
LO-RF
are enabled. When
Frequency Offset
is
ON
,
Base
,
Stimulus
and
Response
are enabled. You can change the
frequency data for each trace.
NOTE
If you use the
LO-RF
setting, you can display information in the same way as with the
mixer measurement function of the Agilent PNA series by changing the measurement data
according to the steps below.
Step 1.
Press
(or
) and
(or
) to activate the trace for
which you want to change the measurement data.
Step 2.
Press
-
Frequency Offset
to display the “Frequency Offset Menu.”
Step 3.
Use the
X-Axis
key to change the frequency data.
Step 4.
Use the
LO-RF
key to change the frequency data to IF frequency.
Step 5.
Convert the measurement data to a complex conjugate number.
Step 6.
Press the
Conversion
key to display the “Conversion Menu.”
Step 7.
Press
Function
to select
Conjugation
.
Step 8.
Press
Conversion
to turn it
ON
.
NOTE
The basic frequency is a frequency range set by the
and
keys in the
STIMULUS bloc.
NOTE
If you change the frequency data, it will affect all of the measurement values, such as the
marker’s read value, conversion parameter, and conversion into time domain.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...