Table 10-3
Header of s4p
!Agilent Technologies,E5071B,<ID>,<FW Revision>
!Date <Date>
!Data & Calibration Information
!Freq Sww:Method(Stat) Sxw:Method(Stat) Syw:Method(Stat) Szw:Method(Stat)
•
•
Swz:Method(Stat) Sxz:Method(Stat) Syz:Method(Stat) Szz:Method(Stat)
# Hz S FMT R Z0
362
Chapter 10
Data Output
Saving Trace Data to a File
Each item has the following meaning.
Sww to Szz
S parameters of the selected test port; corresponds in ascending order,
beginning with 1 to w.
Stat
State of S parameter calibration (ON or OFF)
ON = Calibration is performed.
OFF =Calibration is not performed.
FMT
Data format
RI = Real number- imaginary number
MA = Linear magnitude - angle
DB = Log magnitude - angle
Z0
Reference impedance value
The structure of the data part depends on the combination of the selected file type and
specified port.
NOTE
When the S parameter of a port on which calibration is not performed is specified, if data
measured with the S parameter exists, that data are output to a Touchstone file. If no
measurement data exists, 0 (for log magnitude - angle, log magnitude = -200 dB) is output
in the corresponding field.
Figure 10-4 through Figure 10-7 show the data structures of files saved in Touchstone
format.
Figure 10-4
1-port Touchstone file
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...