Chapter 11
377
Limit Test
Limit Test
11
. Lim
it T
es
t
Figure 11-3
Limit table
Step 6.
Press Add
to add a segment to the limit table and then specify the segment parameter
values shown below.
Segment Parameter
Description
Type
Select the type of segment from the following:
OFF
Segment not used for the limit test
MIN
Segment at which the minimum is specified
MAX
Segment at which the maximum is specified
Begin Stimulus
*1*2
*1.The range in which stimulus values can be specified is from -500 G to +500 G. When a
value outside the range is entered, a suitable value within the range is specified.
*2.Once the stimulus value is specified, changing the sweep range of the E5070B/E5071B
does not affect the stimulus value.
Specify starting point for stimulus value on the limit line
End Stimulus
Specify ending point for stimulus value on the limit line
Begin Response
*3*4
*3.The range in which response values can be specified is from -500 M to +500 M. When a
value outside this range is entered, a suitable value within the range is specified.
*4.After the response value is specified, changing formats results in changing the units but
not the value.
Specify starting point for response value on the limit line
End Response
Specify ending point for response value on the limit line
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...