384
Chapter 11
Limit Test
Ripple Test
Displaying ripple test results
Measurement point and test results
Failed measurement points will be displayed in red on the screen. The test result for the
trace will be indicated as Pass or Fail in the upper-right area of the graph. You can also
display the ripple value at the selected frequency band.
The result will be displayed as Ripl
n
:Pass (or Fail) for each trace.
n
denotes the trace
number. B
n
will be followed by the ripple value (if the ripple display is turned off, only B
n
will be displayed without the ripple value).
For example, in Figure 11-8, “Sample ripple test results,” Ripl1:Pass in the first line
indicates the result for trace 1. The value following B3 is the ripple value at the third
frequency band specified in the ripple test. Similarly, the second line indicates the test
result for trace 2, showing the ripple value at the first frequency band.
For information on how to display the results, see “Turning on/off ripple test and
displaying results” on page 388.
Figure 11-8
Sample ripple test results
Test result for channel
If a channel has a judgment result of “fail,” the message will be displayed as shown in
Figure 11-2. (It will be judged as failed if one or more failed traces are found in the limit
test, ripple test, or bandwidth test within the channel.)
You can also specify the On/Off setting for the Fail Sign in the limit test menu and as well
as in the bandwidth test menu. From the ripple test menu, follow the steps below to turn it
on/off.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...