Chapter 12
403
Optimizing Measurements
Setting Port Extensions and Loss Values
1
2
. O
pt
imiz
ing
M
e
as
ur
em
e
nt
s
2. Setting a DC loss value
Step 1.
Press
or
to activate the channel for which you want to set a DC
loss value.
Step 2.
Press
.
Step 3.
Press
Port Extensions
.
Step 4.
Press
Loss
.
Step 5.
Press
Select Port
to select the port for which you want to set a DC loss value.
Step 6.
Press
Loss at DC
, and enter a DC loss value.
Step 7.
If you want to set a DC loss value for other ports, repeat Step 5 to Step 6.
Enabling port extensions and loss values
Step 1.
Press
or
to activate the channel for which you want to enable
port extensions and loss values.
Step 2.
Press
.
Step 3.
Press
Port Extensions
.
Step 4.
Turn on
Extensions
.