Chapter 12
419
Optimizing Measurements
Performing a Segment-by-Segment Sweep (segment sweep)
1
2
. O
pt
imiz
ing
M
e
as
ur
em
e
nt
s
Procedure
Creating a segment table
Step 1.
Press
or
to select the channel for which you want to create the
segment table.
Step 2.
Press
.
Step 3.
Press
Edit Segment Table
.
The segment table appears in the lower part of the screen.
Step 4.
To change the frequency range setting mode or to set the IF bandwidth, power level, sweep
delay time, sweep mode, and sweep time for each segment, use the following softkeys.
NOTE
When setting the segment table using the front panel keys or keyboard, you need to place
focus on (select) the operation target (segment table of softkey) first. You can change the
focus by pressing
in the ENTRY block. When the focus is placed on the segment
table, the window frame of the segment table is displayed as bright as the window frame of
the active channel. When the focus is placed on the softkey menu, the softkey menu title
area is displayed in blue.
Softkey
Function
Freq Mode
Switches the frequency range setting mode (start/stop or center/span)
List IFBW
Toggles ON/OFF the IF bandwidth setting for each segment; the row for
setting (
IFBW
) only appears in the segment table when this is ON
List Power
Toggles ON/OFF the power level setting for each segment; the row for
setting (
Power
) only appears in the segment table when this is ON
List Delay
Toggles ON/OFF the sweep delay time setting for each segment; the row for
setting (
Delay
) only appears in the segment table when this is ON
List Sweep Mode
Toggles ON/OFF the sweep mode setting for each segment; the row for
setting (
Sweep Mode
) only appears in the segment table when this is ON
List Time
Toggles ON/OFF the sweep time setting for each segment; the row for
setting (
Time
) only appears in the segment table when this is ON
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...