514
Chapter 15
Measurement Examples
Measuring an Unbalanced and Balanced Bandpass Filter
Step 4.
Set the measurement parameter (mixed mode S-parameter) and data format for trace 3.
Setting Description
Key Operation
Unbalanced-balanced conversion of trace 3:
ON
-
-
Fixture Simulator
-
BalUn
(turns it
ON
)
Measurement parameter: S
ss11
-
Sss11
Data format: Smith chart (marker display:
R+jX)
-
Smith
-
R + jX
Step 5.
Set the measurement parameter (mixed mode S-parameter) and data format for trace 4.
Setting Description
Key Operation
Unbalanced-balanced conversion of trace 4:
ON
-
-
Fixture Simulator
-
BalUn
(turns it
ON
)
Measurement parameter: S
dd22
-
Sdd22
Data format: Smith chart (marker display:
R+jX)
-
Smith
-
R + jX
Figure 15-28 shows the setting results for each parameter.
Figure 15-28
Measurement results after unbalanced-balanced conversion