Table 16-3
Corrected System Performance With Type-N Device Connectors, 85092C
Electronic Calibration Module
Network analyzer: E5070B/E5071B, Calibration module: 85092C (Type-N, 50
Ω
) electronic calibration (ECal)
module, Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23
°
C
±
5
°
C with <1
°
C deviation
from calibration temperature, Isolation calibration not omitted
Description
Specification (dB)
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
Directivity
52
52
47
Source Match
45
41
36
Load Match
47
44
39
Reflection Tracking
±
0.040
±
0.060
±
0.070
Transmission Tracking
±
0.039
±
0.069
±
0.136
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
Chapter 16
537
Specifications and Supplemental Information
Corrected System Performance
16. S
pecifica
tions and
Su
pple
m
en
ta
l Infor
m
ation
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...