Chapter 16
559
Specifications and Supplemental Information
Source control
16. S
pecifica
tions and
Su
pple
m
en
ta
l Infor
m
ation
Source control
Measured number of points per sweep
User definable from 2 to 1601.
Sweep mode
Normal stepped, normal swept, fast stepped and fast swept.
Sweep type
Linear sweep, segment sweep, log sweep and power sweep.
Segment sweep
Define independent sweep segments. Set number of points, test port
power levels, IF bandwidth, delay time, sweep time and sweep mode
independently for each segment.
Sweep trigger
Set to continuous, hold, or single, sweep with internal, external, manual,
or bus trigger.
Power
Set source power from -15 dBm (-50 dBm for option 214/314/414) to 10
dBm. The power slope function and the power calibration function
compensate source power level error.
Trace functions
Display data
Display current measurement data, memory data, or current
measurement and memory data simultaneously.
Trace math
Vector addition, subtraction, multiplication or division of measured
complex values and memory data.
Title
Add custom title to each channel window. Titles are printed on
hardcopies of displayed measurements.
Autoscale
Automatically selects scale resolution and reference value to vertically
center the trace.
Electrical delay
Offset measured phase or group delay by a defined amount of electrical
delay, in seconds.
Phase Offset
Offset measured phase or group delay by a defined amount in degrees.
Statistics
Calculates and displays mean, standard deviation and peak-to-peak
deviation of the data trace.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...