Calibration Type
Measurement ports of
Reflection and Match cal.
Measurement paths of Thru
and Line cal.
3-port TRL/LRM cal. for
Ports: 1-2-3
All ports (1, 2, and 3)
1-2 and 1-3
3-port TRL/LRM cal. for
Ports: 1-2-4
All ports (1, 2, and 4)
1-2 and 1-4
3-port TRL/LRM cal. for
Ports: 1-3-4
All ports (1, 3, and 4)
1-3 and 3-4
3-port TRL/LRM cal. for
Ports: 2-3-4
All ports (2, 3, and 4)
2-3 and 3-4
4-port TRL/LRM cal. for
Ports: 1-2-3-4
All ports (1, 2, 3, and 4)
1-2, 1-3, and 3-4
Appendix A
593
Manual Changes
Manual Changes
A. Manual
Changes
Step 1.
Measure calibration data.
In this example, 3-port TRL/LRM calibration for test ports 1, 2, and 3 (1 in Figure A-8) is
performed by using the Match, Line 1, and Line 2 standards.
Figure A-8
Performing multiport TRL/LRM calibration
Step 2.
Press the
Thru
button (2 in Figure A-8).
Step 3.
Make the through condition for test ports 1 and 2 and press the
[ ]Thru 1-2
button (1 in
Figure A-9). The through measurement is performed, and an asterisk appears at the
[ ]Thru
1-2
button.
Step 4.
Make the through condition for test ports 1 and 3, and press the
[ ]Thru 1-3
button (2 in
Figure A-9). The through measurement is performed, and an asterisk appears at the
[ ]Thru
1-3
button and then a check mark [v] (3 in Figure A-9) also appears on the
Thru
button.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...