Appendix A
603
Manual Changes
Manual Changes
A. Manual
Changes
*1.The test port dynamic range is calculated as the difference between the test port rms noise floor and the source
maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals
into account.
*2.May be limited to 90 dB at particular frequencies below 350MHz or above 4.25GHz due to spurious receiver
residuals.
Table A-5
Test Port Output Power
*1
Description
Specification
Supplemental Information
Level Accuracy (at 23
°
C
±
5
°
C)
300 kHz to 10 MHz
10 MHz to 8.5 GHz
±
0.650 dB (at 0 dBm, 50 MHz
absolute)
±
1.0 dB (at 0 dBm, relative to 50
MHz reference)
±
1.0 dB (at 0 dBm, relative to 50
MHz reference)
Level Accuracy (high temperature mode: ON)
300 kHz to 8.5 GHz
±
0.8 dB (at 0 dBm, 50 MHz
absolute)
±
1.5 dB (at 0 dBm, relative to 50
MHz reference)
Level Accuracy (swept mode)
300 kHz to 4.25 GHz
4.25 GHz to 8.5 GHz
±
2.5 dB (at 0 dBm, relative to 50
MHz reference)
±
3.5 dB (at 0 dBm, relative to 50
MHz reference)
Level Linearity (23
°
C
±
5
°
C)
10 MHz to 4.25 GHz
4.25 GHz to 6 GHz
6 GHz to 8.5 GHz
±
0.75 dB (at
−
15 dBm to 10 dBm)
±
0.75 dB (at
−
15 dBm to 8 dBm)
±
0.75 dB (at
−
15 dBm to 6 dBm)
Level Linearity (high temperature mode: ON)
300 kHz to 4.25 GHz
4.25 GHz to 6 GHz
6 GHz to 8.5 GHz
±
1.5 dB (at
−
15 dBm to 10 dBm)
±
2.0 dB (at
−
15 dBm to 8 dBm)
±
2.0 dB (at
−
15 dBm to 6 dBm)
Level Linearity (swept mode)
300 kHz to 4.25 GHz
4.25 GHz to 6 GHz
6 GHz to 8.5 GHz
±
1.5 dB (at
−
15 dBm to 10 dBm)
±
3 dB (at
−
15 dBm to 8 dBm)
±
3 dB (at
−
15 dBm to 6 dBm)
Range
Standard
300 kHz to 4.25 GHz
4.25 GHz to 6 GHz
6 GHz to 8.5 GHz
Extended Power Range (with option 214, 314,
414)
300 kHz to 4.25 GHz
4.25 GHz to 6 GHz
6 GHz to 8.5 GHz
−
15 dBm to 10 dBm
−
15 dBm to 8 dBm
−
15 dBm to 6 dBm
−
50 dBm to 10 dBm
−
50 dBm to 8 dBm
−
50 dBm to 6 dBm
(non-harmonics spurious may
limit power range)
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...