Change 4
The firmware revision A.03.54 or below does not support the following functions. Please
delete the descriptions about these functions in this manual.
o Auto Power Range set function
Change 3
The equipments that firmware revision at the factroy shipment is A.03.53 or below is not
eligible for the following specs. support the following functions. Please refer to original
attached manuals about them.
o Initial Source Port Control function
Change 2
Figure 16-4 dimensions of rear view was changed. Eligible serial numbers are
MY42100211 (for E5070B), MY42100407 (for E5071B) and later. For the earlier serial
number equipment, please refer to original attached manuals.
Change 1
The firmware revision A.03.0x does not support the following functions. Please delete the
descriptions about these functions in this manual
o User Characterized ECal
o Confidence check on calibration coefficients using ECal
o 4-port network embedding/de-embedding (Fixture simulator)
300 kHz to 3 MHz
(source power level = +10 dBm)
0.035
°
rms (23
°
C
±
5
°
C, typical)
0.05
°
rms (high temperature
mode: ON, typical)
3 MHz to 4.25 GHz
(source power level = +10 dBm)
0.007
°
rms (23
°
C
±
5
°
C, typical)
0.02
°
rms (high temperature
mode: ON, typical)
4.25 GHz to 6 GHz
(source power level = +8 dBm)
0.021
°
rms (23
°
C
±
5
°
C, typical)
0.035
°
rms (high temperature
mode: ON, typical)
6 GHz to 7.5 GHz
(source power level = +6 dBm)
0.021
°
rms (23
°
C
±
5
°
C, typical)
0.035
°
rms (high temperature
mode: ON, typical)
7.5 GHz to 8.5 GHz
(source power level = +6 dBm)
0.035
°
rms (23
°
C
±
5
°
C, typical)
0.05
°
rms (high temperature
mode: ON, typical)
*1.Trace noise is defined as a ratio measurement of a through, at IFBW = 3 kHz.
Table A-7
Test Port Input (Trace Noise)
Description
Specification
Supplemental Information
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...