Chapter 3
71
Setting Measurement Conditions
Setting Stimulus Conditions
3
. Set
ting
M
e
as
ur
em
e
nt
C
onditions
Setting the Sweep Range with the Lowest and Highest Values
Step 1.
Press
or
to select the channel whose sweep range will be set.
Step 2.
Press
.
Step 3.
Using the ENTRY block keys on the front panel, input the lowest value.
Step 4.
Press
.
Step 5.
Using the ENTRY block keys on the front panel, input the highest value.
Setting the Sweep Range with the Center Value and a Span
Step 1.
Press
or
to select the channel whose sweep range will be set.
Step 2.
Press
.
Step 3.
Using the ENTRY block keys on the front panel, input the center value.
Step 4.
Press
.
Step 5.
Using the ENTRY block keys on the front panel, input a span value.
Setting the Sweep Range Using the Marker
Step 1.
In the channel window whose range must be set, place the active marker on the active trace
to a position that corresponds to the new range (to the lowest, highest, or center value).
Step 2.
Press
.
Step 3.
Press the softkey that corresponds to each value.
Softkey
Function
Marker
→
Start
Sets the lowest value to the stimulus value of the active marker on the
currently active trace.
Marker
→
Stop
Sets the highest value to the stimulus value of the active marker on the
currently active trace.
Marker
→
Center
Sets the center value to the stimulus value of the active marker on the
currently active trace.
NOTE
If the reference marker is on and the stimulus value of the active marker is expressed by a
value relative to the reference marker, the absolute stimulus value will be used to set the
new sweep range.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...