Chapter 4
113
Calibration
1-Port Calibration (reflection test)
4. Cal
ibra
tion
1-Port Calibration (reflection test)
In 1-port calibration, calibration data are measured by connecting an OPEN standard, a
SHORT standard, and a LOAD standard to the desired test port. This calibration effectively
eliminates the frequency response reflection tracking error, directivity error, and source
match error from the test setup in a reflection test using that port (Figure 4-11).
Figure 4-11
1-Port error model (1-port calibration)
Figure 4-12
Connecting the standard for 1-port calibration