130
Chapter 4
Calibration
ECal (electronic calibration)
Unknown Thru Calibration
E5070B/E5071B allows you to perform thru calibration of ECal as unknown thru
calibration. In this function, the thru calibraion is done with thru standard in ECal,
however, the stored thru calibration data in ECal is not used. E5070B/E5071B performs
thru calibration as a unknown thru calibration.
For more detail on the unknown thru calibration, see “Unknown Thru Calibration” on
page 153.
Step 1.
Press
.
Step 2.
Press
ECal
.
Step 3.
Press
Unknown Thru
to turn on.
Step 4.
Perform your desired ECal calibration.