Calibration
Changing the Calibration Kit Definition
Step 4.
Press
Modify Kit
. To change the pre-defined connector type (e.g. OPEN(f) to OPEN (m))
skip to Step 14.
Step 5.
Press
Define STDs
.
Step 6.
Select the standard to be redefined from among standards numbered 1 through 21.
Step 7.
Press
STD Type
.
Step 8.
Select the type of standard according to the list below.
Softkey
Function
Open
Selects the OPEN standard
Short
Selects the SHORT standard
Load
Selects the LOAD standard
Delay/Thru
Selects the delay/THRU standard
Arbitrary
Selects the arbitrary impedance
None
Selects no standard type
Step 9.
Set the standard coefficient according to the list below.
Softkey
Function
C0
Sets C0
C1
Sets C1
C2
Sets C2
C3
Sets C3
L0
Sets L0
L1
Sets L1
L2
Sets L2
L3
Sets L3
Offset Delay
Sets the offset delay
Offset Z0
Sets the offset Z0
*1.Specify 1
Ω
when Waveguide is set for the media type.
*1
Offset Loss
Sets the offset loss
Arb. Impedance
Sets an arbitrary impedance
Min. Frequency
Sets a start frequency.
*2.Specify cutoff frequency when Waveguide is set for the media type.
*2
Max. Frequency
Sets a stop frequency.
Connector Type
Selects a media type.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...