202
Chapter 4
Calibration
Specifying Different Standard for Each Frequency
Specifying Different Standard for Each Frequency
Defining different standard for each frequency band
This section demonstrates the procedure to define a different open standard for each
frequency band, based on the following information.
Table 4-6
Example of standard definition
Standard
Label name
Frequencies to be defined
Open
Open 3G
1 GHz - 3 GHz
Open
Open 6G
3 GHz - 6 GHz
Procedure
Step 1.
Press
.
Step 2.
Press
Cal Kit
.
Step 3.
Select a calibration kit you want to define.
Step 4.
Press
Modify Kit
.
Step 5.
Press
Define STDs
.
Step 6.
Select #1, press
Label
, and enter
Open 3G
using the character input pad that appears on the
screen.
Step 7.
Select
Open
as the standard type (
STD Type
).
Step 8.
Set a necessary standard coefficient.
Step 9.
Enter
1G
as the minimum frequency for
Min Frequency
.
Step 10.
Enter
3G
as the maximum frequency for
Max Frequency
.
Step 11.
Press
Return
.
Step 12.
In the same way, select #2, press
Label
, and enter
Open 6G
using the character input pad
that appears on the screen.
Step 13.
Select
Open
as the standard type (
STD Type
).
Step 14.
Set a necessary standard coefficient.
Step 15.
Enter
3G
as the minimum frequency for
Min Frequency
.
Step 16.
Enter
6G
as the maximum frequency for
Max Frequency
.
Step 17.
Press
Return
.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...