Chapter 8
321
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
8. Fre
q
uency
Of
fset
Me
asur
eme
nt
(Option 0
08)
Figure 8-8
Avoid Spurious Function
NOTE
You cannot avoid all spurious signals by using the avoid spurious function.
Procedures for controlling spurious avoidance function
Follow the steps below to control the spurious avoidance function.
Step 1.
Press
-
Frequency Offset
to display the “Frequency Offset Menu.”
Step 2.
Press
Avoid Spurious
to turn it ON.
NOTE
Even if the frequency-offset function is turned
OFF
, the function is available when the
external signal source (
LO Frequency
) is
ON
.
NOTE
If the
Swept
is selected for the
Sweep Mode
, it will be changed to
Stepped
(from
Std
Swept
to
Std Stepped
, and from
Fast Swept
to
Fast Stepped
).
4. Changing Frequency Data
The frequency-offset function offers the capability to change the frequency data of each
trace to any desired frequency.
Procedures for changing frequency data
Step 1.
Press
(or
) and
(or
) to activate the trace for
which you want to change the frequency.
Step 2.
Press
-
Frequency Offset
to display the “Frequency Offset Menu.”
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...