Chapter 10
355
Data Output
Saving and Recalling Instrument State
10. Da
ta Ou
tp
ut
Saving Instrument State
Follow the procedure below to save internal data from the E5070B/E5071B.
Step 1.
Press
.
Step 2.
Press
Save State
.
Step 3. When you want to use a pre-defined file name (State01.sta - State08.sta,
Autorec.sta)
Press
State01
-
State08
or
Autorec
.
NOTE
If “A:\Autorec.sta” or “D:\Autorec.sta” is found on the system at startup, the
E5070B/E5071B is automatically configured using the saved settings. When both files are
found, “A:\Autorec.sta” is recalled. To disable the auto recall function, delete the
Autorec.sta files.
NOTE
An asterisk (*) in the upper right of the softkey indicates that the corresponding file of the
softkey already exists. If you save into the existing file, the existing file is copied as
“backup.sta” and then overwritten.
When you want to use another file name or a file name on the floppy disk
1. Press
File Dialog...
to open the Save As dialog box. Figure 10-1 illustrates the Save As
dialog box.
2. Select the folder and input a file name using the external keyboard and mouse.
3. Click
Save
Figure 10-1
Save As dialog box
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...