376
Chapter 11
Limit Test
Limit Test
Defining the limit line
To use the limit test, you must first define the limit line. You can define a limit table for
each trace, and you can define up to 100 limit lines (segments) in a limit table.
Defining a segment
The following steps describe how to define a segment.
Step 1.
Press
or
to activate the channel on which the limit test function
will be used.
Step 2.
Press
or
to activate the trace on which the limit test function will be
used.
Step 3.
Press
to display the Analysis menu.
Step 4.
Press
Limit Test
to display the softkeys associated with the limit test.
Step 5.
Press
Edit Limit Line
to display the limit table shown in Figure 11-3. Using the limit table,
create/edit a segment. Initially, no segments are entered in the limit table.
At the same time, the Edit Limit Line menu used to create/edit the limit table is displayed.
The following lists the functions that correspond to the softkeys shown.
Softkey
Function
Delete
Deletes the segment containing the selected cell from the limit table.
Add
Adds a segment before the segment containing the selected cell to the
limit table.
Clear Limit Table
Clears all of the limit table data.
Export to CSV File
Exports the limit table to a file in CSV (comma-separated value)
format. For more information, see “Saving/calling the limit table” on
page 379.
Import from CSV File
Imports a limit table saved in CSV (separated-separated value)
format. For more information, see “Saving/calling the limit table” on
page 379.
Return
Exits the Edit Limit Line menu and closes the limit table display.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...