Limit Test
Ripple Test
Figure 11-9
Ripple limit table
Step 6.
Press
Add
to add a frequency band to the ripple limit table and then specify the following
parameters for the frequency band:
Parameter
Description
Type
Selects a frequency type, either ON or OFF.
ON
Band used for the ripple test.
OFF
Band not used for the ripple test.
Begin Stimulus
*1*2
*1. Acceptable range for the stimulus value: -500G to +500G. If any outranging value is
specified, it will be reset to fall within the range.
*2. Even if the E5070B/E5071B's sweep range is changed after the stimulus value has been
set, the stimulus value is not susceptible.
Specifies the start point for the stimulus value in the ripple test.
End Stimulus
Specifies the end point for the stimulus value in the ripple test.
Max Ripple
Specifies the ripple limit value.