496
Chapter 15
Measurement Examples
Evaluating a Duplexer
Figure 15-9
EcalAssistant (port/channel selection) dialog box
Step 5.
Following the instructions in the dialog box, select the type of ECal, test ports, and the
channel.
Setup Description
Key Operation
Select Ports
ECal type: Full 3-port calibration
3 Port
Test ports to be used for ECal: 1, 2, and 3
1, 2, 3
Select Channel
Channel on which ECal is to be executed:
Channel 1
Channel
:
1
Step 6.
Press the
Next
button. The EcalAssistant (connection) dialog box (Figure 15-10) appears.
Figure 15-10
EcalAssistant (connection) dialog box
Step 7.
Connect the ECal module to the test ports according to the instructions in the dialog box
and then press
Measure
.
Setup Description
Key Operation
Executing calibration between ports 1 and 2
(After making the connections shown in Figure
15-11 (a)), press
Measure
-
Next
Executing calibration between ports 1 and 3
(After making the connections shown in Figure
15-11 (b)), press
Measure
-
Next
Executing calibration between ports 2 and 3
(After making the connections shown in Figure
15-11 (c)), press
Measure
-
Next
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...