Chapter 15
505
Measurement Examples
Measuring the Deviation from a Linear Phase
15
. Me
asur
eme
n
t Exa
m
ple
s
Measuring the Deviation from a Linear Phase
This section illustrates how to determine the deviation from a linear phase in the pass band
of a 1.09-GHz bandpass filter.
Evaluation Steps
Here, the DUT is evaluated by following the steps described in Table 15-5.
Table 15-5
Evaluating the Deviation from a Linear Phase
Step
Description
“1. Connect the DUT” on page 505
The DUT is connected.
“2. Define the Measurement Conditions” on
page 506
The measurement conditions are defined.
“3. Execute the Calibration” on page 506
The calibration is executed.
“4. Connect the DUT and Execute the Auto
Scale” on page 506
The DUT is connected again to execute the auto scale
function.
“5. Specify the Electrical Delay” on page 507
The electrical delay is specified.
“6. Measure the Deviation from a Linear
Phase” on page 508
The statistics data function (peak-to-peak) is used to
determine the deviation from a linear phase.
1. Connect the DUT
Connect the DUT as shown in Figure 15-21.
Figure 15-21
Connecting the DUT
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...