Chapter 15
513
Measurement Examples
Measuring an Unbalanced and Balanced Bandpass Filter
15
. Me
asur
eme
n
t Exa
m
ple
s
4. Setting a Balance Conversion Topology
Follow the procedure below to set the balanced conversion topology.
Table 15-7
Setting Description
Key Operation
Set port 1 on the DUT to unbalanced and port 2
on the DUT to balanced.
-
Fixture Simulator
-
Topology
-
Device
-
SE-Bal
(check only)
Set the connecting destination of port 1 on the
DUT (unbalanced) to test port 1 of the
analyzer.
Port 1 (se)
-
1
(check only)
Set the connecting destination of port 2 on the
DUT (balanced) to test ports 2 and 3 of the
analyzer.
Port 2 (bal)
-
2-3
(check only)
5. Selecting Measurement Parameters
Step 1.
Display four traces.
Setting Description
Key Operation
Number of traces: 4
-
Number of Traces
-
4
Trace allocation: 4-part split
Allocate Traces
-
Step 2.
Set the measurement parameter (mixed mode S-parameter) and data format for trace 1.
Setting Description
Key Operation
Fixture simulator: ON
-
Fixture Simulator
-
Fixture Simulator
(turns it
ON
)
Unbalanced-balanced conversion of trace 1:
ON
BalUn
(turns it
ON
)
Measurement parameter: S
ds21
-
Sds21
Step 3.
Set the measurement parameter (mixed mode S-parameter) and data format for trace 2.
Setting Description
Key Operation
Unbalanced-balanced conversion of trace 2:
ON
-
-
Fixture Simulator
-
BalUn
(turns it
ON
)
Measurement parameter: S
cs21
-
Scs21
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...