Table 16-11
Test Port Input (Trace Noise)
Description
Specification
Supplemental Information
Trace Noise
*1.Trace noise is defined as a ratio measurement of a through, at IFBW = 3 kHz.
*1
Magnitude
300 kHz to 3 MHz
(source power level = +10 dBm)
5 mdB rms (typical)
8 mdB rms (high temperature
mode: ON, typical)
3 MHz to 3 GHz
(source power level = +10 dBm)
1 mdB rms (23
°
C
±
5
°
C)
4 mdB rms (high temperature
mode: ON, typical)
3 GHz to 4.25 GHz
(source power level = +9 dBm)
1.2 mdB rms (23
°
C
±
5
°
C)
4.8 mdB rms (high temperature
mode: ON, typical)
4.25 GHz to 6 GHz
(source power level = +7 dBm)
3.6 mdB rms (23
°
C
±
5
°
C)
7.2 mdB rms (high temperature
mode: ON, typical)
6 GHz to 7.5 GHz
(source power level = +5 dBm)
3.6 mdB rms (23
°
C
±
5
°
C)
7.2 mdB rms (high temperature
mode: ON, typical)
7.5 GHz to 8.5 GHz
(source power level = +5 dBm)
6 mdB rms (23
°
C
±
5
°
C)
9.6 mdB rms (high temperature
mode: ON, typical)
Trace Noise
Phase
300 kHz to 3 MHz
(source power level = +10 dBm)
0.035
°
rms (23
°
C
±
5
°
C, typical)
0.05
°
rms (high temperature
mode: ON, typical)
3 MHz to 3 GHz
(source power level = +10 dBm)
0.007
°
rms (23
°
C
±
5
°
C, typical)
0.02
°
rms (high temperature
mode: ON, typical)
3 GHz to 4.25 GHz
(source power level = +9 dBm)
0.008
°
rms (23
°
C
±
5
°
C, typical)
0.024
°
rms (high temperature
mode: ON, typical)
4.25 GHz to 6 GHz
(source power level = +7 dBm)
0.025
°
rms (23
°
C
±
5
°
C, typical)
0.042
°
rms (high temperature
mode: ON, typical)
6 GHz to 7.5 GHz
(source power level = +5 dBm)
0.025
°
rms (23
°
C
±
5
°
C, typical)
0.042
°
rms (high temperature
mode: ON, typical)
7.5 GHz to 8.5 GHz
(source power level = +5 dBm)
0.042
°
rms (23
°
C
±
5
°
C, typical)
0.06
°
rms (high temperature
mode: ON, typical)
544
Chapter 16
Specifications and Supplemental Information
Test Port Input
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...