Table 16-12
Test Port Input (Stability)
Description
Specification
Supplemental Information
Stability Magnitude
*1.Stability is defined as a ratio measurement at the test port.
*1
3 MHz to 3 GHz
0.005 dB/
°
C
(at 23
°
C
±
5
°
C, typical)
3 GHz to 6 GHz
0.01 dB/
°
C
(at 23
°
C
±
5
°
C, typical)
6 GHz to 8.5 GHz
0.04 dB/
°
C
(at 23
°
C
±
5
°
C, typical)
Stability Phase
3 MHz to 3 GHz
0.1
°
/
°
C
(at 23
°
C
±
5
°
C, typical)
3 GHz to 6 GHz
0.2
°
/
°
C
(at 23
°
C
±
5
°
C, typical)
6 GHz to 8.5 GHz
0.8
°
/
°
C
(at 23
°
C
±
5
°
C, typical)
Table 16-13
Test Port Input (Dynamic Accuracy)
Accuracy of the test port input power reading is relative to
−
10 dBm reference input power level.
Specification
Supplemental Information
Chapter 16
545
Specifications and Supplemental Information
Test Port Input
16. S
pecifica
tions and
Su
pple
m
en
ta
l Infor
m
ation
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...