Table 16-14
Test Port Input (Group Delay)
*1
Description
Specification
Supplemental Information
Aperture (selectable)
(frequency span)/(number of points
−
1)
Maximum Aperture
25% of frequency span
Minimum Delay
Limited to measuring no more
than 180
°
of phase change within
the minimum aperture.
Accuracy
See graph below, typical
The following graph shows group delay accuracy with type-N full 2-port calibration and a 10 Hz IF bandwidth. Insertion
loss is assumed to be < 2 dB.
In general, the following formula can be used to determine the accuracy, in seconds, of specific group delay measurement:
±
Phase Accuracy (deg) / [360
×
Aperture (Hz)]
546
Chapter 16
Specifications and Supplemental Information
Test Port Input
*1.Group delay is computed by measuring the phase change within a specified step (determined by the frequency
span and the number of points per sweep).