Appendix A
587
Manual Changes
Manual Changes
A. Manual
Changes
Figure A-2
Turning off system error correction
NOTE
The System Correction Off button does not appear when system error correction has
already been turned off.
2-port TRL/LRM calibration operating procedure
1. Selecting Test Port
Select test ports (1 in Figure A-3).
NOTE
For more information on starting VBA macro, see “Operating Procedure” on page 586
.
Figure A-3
TRL/LRM main control panel
NOTE
The TRL/LRM calibration VBA macro automatically detects the ENA’s active channel,
and calibration is performed for the active channel.
2. Defining Calibration Kit
Define the calibration kit you use.
Step 1.
Press the
Define Cal kit
button (2 in Figure A-3) to bring up the calibration kit definition
screen shown in Figure A-4.